Today, let me talk about the common problems and troubleshooting methods of home oxygen absorbers. The following is a brief list of common problems and troubleshooting methods in several categories: http://zhiyangjidi.com/
1. Fault phenomenon: there is gas in the airflow
Cause Analysis:
1. In the initial stage of use, the connecting pipe and other plastic parts will emit gas. This phenomenon is normal.
Remedy: This gas is the normal smell of silicone tube and ABS plastic. It has been tested to be non-toxic and harmless. With the use of the machine, the gas will slowly disappear.
2. The wetted water tank has not been cleaned for a long time, or the water has not been changed for a long time, resulting in odor in the water tank.
Remedy: Clean the oxygen pipe and the humidification tank.
2. Fault phenomenon: water is emitted from the oxygen outlet. Http://zhiyangjidi.com/article/122.html
Cause Analysis:
1. The water in the humidification tank exceeds the highest water level line, and the air flow presses the water bubbles into the gas pipeline.
Remedy: Reduce the water level in the humidification tank below the maximum water level.
2. Long-term use of water vapor in the machine airflow to condense on the pipe wall
Remedy: Pour out the water in the water tank, wait until no water emerges from the oxygen outlet, and then add water to the water tank.
3. Fault phenomenon: the machine emits a '呲' sound every few seconds.
Analysis of the cause: This is the nitrogen discharge sound when the machine is working normally, which is a sign that the machine is working well.
4. Fault phenomenon: After starting the machine, the machine is running normally, the flow meter shows '0' or is too small
Cause Analysis:
1. The flow meter is off or the value is too small
Remedy: Increase the flowmeter
2. The oxygen pipe is kinked or blocked
Remedy: rationalize the oxygen pipe
3. The internal connecting pipe of the machine is detached or broken
Remedy: Cut off the power supply and repair by professional maintenance personnel.
V. Fault phenomenon: the water tank bubbles when the machine is turned on, and the oxygen outlet has no oxygen.
Analysis of the cause: the humidified water tank rubber plug is not plugged
Remedy: Remove the humidification tank and plug the rubber stopper
VI. Fault phenomenon: The indicator light is normal after booting, the machine is not working normally [Working sound is not normal]
Cause Analysis:
Compressor failure
Remedy: Turn off the power and perform maintenance by professional maintenance personnel.
2. The ambient temperature is too high, and the self-protection function of the oil-free compressor in the machine starts.
Remedy: For normal phenomenon, the machine will restart automatically after the temperature drops.
3. Separation valve failure
Remedy: cut off the power supply by professional maintenance personnel
4. The internal connecting pipe of the machine is detached or broken
Remedy: Cut off the power supply by professional maintenance personnel. The above are common problems and troubleshooting methods for household oxygen absorbers. I hope this article is helpful to everyone.